Scanner 11 Nov: hot-fixes from the 7 T session - #232
Merged
Merged
Conversation
The z lobes of BssfpReadout2D/3D were all built on the full-scale ramp max_grad / max_slew, so that a slice rephaser and a partition encode sum to one trapezoid. At 80 mT/m and 132.6 T/m/s that ramp is 604 us, and a small 2D slice rephaser became a 1.208 ms triangle at 19 T/m/s that set both windows of the repetition (TR 5.016 ms at 7 T for 192x192, 300 mm, 8 mm, 125 kHz). The shared ramp is now the one on which the largest summed lobe is fastest at the slew rate, the half-flip lobe has its own, and each window takes only its own floors. Same protocol: 3.736 ms at 132.6 T/m/s; 3.912 ms at the 105 T/m/s the design server lowers it to for PNS. Balance is exact (moments below 5e-14 of a k-space step on every axis). An ungated bssfp2D_sequence now plays each slice's lines n_phases times back to back, a free-running series labelled PHS, without dummies between frames; it used to ignore n_phases and play one frame. Co-Authored-By: Claude Opus 5.5 <[email protected]>
… flip The gradient-echo calibration that epi2d plays when undersampled or multiband, and epi3d when undersampled, excited every line with the imaging pulse at the imaging flip. Its lines follow one another at its shortest TR, so at 7 T a 70 degree scan (exam 28827: 2 slices, R 3, calibration TR 6.4 ms) was refused at predownload: "SAR at 117% of its 10 s limit: 6.4 ms of scan need 7.5 ms". At that flip the calibration also read its k-space centre at about a hundredth of the first line's signal. The calibration now plays CALIBRATION_FLIP_DEG, 10 degrees, near the Ernst angle of its TR: (10/70)^2 of the RF power, about 2 % of the limit in that scan, and its signal held across the calibration region. The calibration only measures coil sensitivities. epi2d scales the single-band pulse's samples (a small-tip SLR pulse, no second design or bandwidth measurement); epi3d designs its excitation at the imaging pulse's duration. Either way the calibration plays the imaging waveform, so the RF definition and the segmentation are unchanged. Co-Authored-By: Claude Opus 5.5 <[email protected]>
…led read axis On the 7 T (11 Nov, series 15-17) the two forward lines of the three-line navigator differed by a linear phase of 2.0-2.35 rad, an echo shift of 0.64 samples: the first lobe after the prewinder echoes away from where the steady train does. The odd-even phase fitted to the navigator was then off by about 0.5 rad and 0.8-1.4 rad/half-FOV, leaving an N/2 ghost of 15 % in a single-shot EPI where the best linear correction leaves 6 %. The train now plays one negative lobe, without ADC, between the read prewinder (which opposes it) and the navigator; the navigator keeps its + - + polarity and its labels, and the train its alternation. The echo time of the centre line grows by one echo spacing at the shortest TE; the spacing, the shapes and the ADC are unchanged (one more block per shot, a scaled copy of the read lobe). Co-Authored-By: Claude Opus 5.5 <[email protected]>
A scanner that prepares a segment between blocks at rest cut a slice of the balanced SSFP into one segment a repetition, and took its segment time after each: the played repetition was longer than the designed one and the echo sat ahead of its middle. BssfpReadout2D(bridge=True) runs the ramps of the selection plateau through zero to a fraction of its amplitude where the excitation block meets the rewind before it and the acquisition after it, and the rephasers end and start there. No join between two repetitions of a slice then has every gradient at rest, the areas either side of the pulse centre are unwound exactly, and the block durations are those of a slice axis that rests; where they could not be, the slice is left unbridged (bridged is False). gz_open opens a slice from rest and gz_close closes its last rewind to rest. bssfp2d bridges the slices of a free-running scan of several slices. A gated train rests where it waits for the heartbeat. A single slice is no repetition of the scan, and a receive calibration then plays one repetition alone, from rest to rest, which a bridged repetition is not. The 3D readout is unchanged. Co-Authored-By: Claude Fable 5.2 <[email protected]>
One slice is the several-slice scan with one slice: it opens and closes at rest and is played as one segment. The receive calibration of a single bridged slice is the scan's own start (pulserver). Co-Authored-By: Claude Opus 5.5 <[email protected]>
A gated slice rested and waited for a trigger at every heartbeat, so its repetitions could not be joined and a scanner played them a segment each. Every 2D slice is now bridged. Under prospective gating a slice waits for one trigger, ahead of its half flip, and then runs free at the nominal heart rate: each heartbeat plays its segment once per cardiac phase and repetitions that do not acquire fill the rest of it (prospective_starts), so the steady state is kept across heartbeats. A heart rate away from the nominal one drifts against the cardiac phases, which a trigger per heartbeat corrected. Retrospective gating plays as before, with no trigger, and is bridged. Co-Authored-By: Claude Opus 5.5 <[email protected]>
…d raster The bridged readout fell back to a slice axis that rests wherever a bridged rephaser was longer than the window a resting one needs. On a 4 us gradient raster that is every prescription, by a raster step, so a scanner played an unbridged slice at any repetition time, the minimum and a longer one alike. bridge=True now always bridges. The windows either side of the excitation hold the bridged rephasers; the shortest repetition is then within a few rasters of that of a resting slice axis (the same at 105 T/m/s and above on both rasters, up to 0.24 ms longer at 30 T/m/s). A longer repetition time lengthens the acquisition and rewind blocks, which carry the read and phase lobes from one end to the other, so no block and no join rests. The half-flip gap holds its two rephasers apart where it has the room, and one lobe held at the bridge level where it does not. gz_close is None where the closing lobe does not fit the last rewind. Co-Authored-By: Claude Opus 5.5 <[email protected]>
…eiling The spoilers of a fat saturation or a saturation band play on three axes at once. Designed at the sequence's slew they set its nerve stimulation, and the design server's slew search then slows every gradient of the sequence for them: a 2D EPI with fat saturation was designed at 49.9 T/m/s instead of 88, the spoiler alone at 0.95 of the stimulation limit and the train at 0.60. epi2D, gre2D and se2D now build these modules under pp.cap_system(system, max_slew=PREPARATION_MAX_SLEW), 40 T/m/s; the spoiler of the fat saturation grows from 2.76 to 2.88 ms and the train keeps the sequence's slew. The safety search still checks the whole sequence. Co-Authored-By: Claude Opus 5.5 <[email protected]>
…ities The calibration of an undersampled 2D or 3D EPI reads its central lines in n_shots * ry interleaved EPI shots with the imaging train's read lobe, navigator, blip window, echo spacing and echo time (Polimeni et al., MRM 2016), a slice's or partition's shots in rapid succession at 10 degrees, forward (SET 0) and reversed (SET 1). It is the default; calibration='gre' keeps the gradient echo per line. EpiReadout2D/3D take blip_span, so the calibration matches a multiband train's echo spacing. Co-Authored-By: Claude Opus 5.5 <[email protected]>
Co-Authored-By: Claude Opus 5.5 <[email protected]>
The trigger had a block of its own ahead of the half flip, a segment the scanner played for microseconds, and the slice's train after it had to be set up in that time: a slice-long train cannot be, and the interpreter refuses it. The trigger now sits on the half-flip block, so the train itself waits for it and is set up before the trigger, or while the slice before it plays. The trigger delay is played as repetitions that do not acquire (prospective_starts), keeping the steady state through it. Co-Authored-By: Claude Opus 5.5 <[email protected]>
The reference (reversed phase encode) and the gradient-echo calibration stated no kSpaceCenterLine (3D: kSpaceCenterPartition), so pulserver's enrichment centred them on the lines they read. Under partial Fourier the reference's lines 2..144 of 192 were placed 23 rows off (exam 5190 series 32), which misplaces its echo for the homodyne completion and for any joint reconstruction with the forward shots. Co-Authored-By: Claude Opus 5.5 <[email protected]>
… frames Matteo, 12 Nov: the frames of a 2D cine must not be one segment. An ungated slice with dummies is now two files of the chain: the half flip and dummies, closed to rest, then the frames, each opened from rest (gz_open) and closed to it (gz_close) at the bridged repetition time, the file's TRSize one frame. A scanner that segments a repetition then prepares one segment for the dummies and one for a frame, played once per frame, and dedups both across slices. Gated slices are one train as before. BssfpReadout2D(bridge=True): gz_close is the rest-to-rest lobe that rephases gz_open, always present, and the rewind window holds it (the repetition time at 7 T limits is unchanged at 192 and 256). Co-Authored-By: Claude Opus 5.5 <[email protected]>
…st, a FIESTA-like pulse A free-running slice is no longer bridged: every repetition opens from rest and closes to it, and its files declare no TRSize, so a scanner that segments a repetition plays each repetition as the same segment and adds its segment start time to every repetition time alike. A frame-long segment lengthened only the repetition at each frame boundary and set off a transient in the steady state there. Gated slices stay bridged, one train with the designed timing against the heartbeat. The half flip is replaced by a linear ramp of the flip angle over the first CATALYST_RAMP (20) dummies. It damps the oscillating transient better off resonance (worst over +-0.45/TR at TR 1: 9 % against 46 %) and frees the repetition from the half-flip timing, which refused 166-250 kHz at 192^2. Repetitions are counted from the slice's first excitation, the trigger's under prospective gating, with no half-repetition offset. Default dummies bring the steady state within STEADY_STATE_TOLERANCE (5 %), and the selective pulse defaults to 0.48 ms at a time-bandwidth product of 2, as product balanced SSFP excites (about half the RF energy of 0.6 ms at 4), both pulse settings now arguments; the default bandwidth is 250 kHz. The paper-plot and expand-repeats tests, which the chain commit had broken (a chain given to the plot; a frames file without ONCE), now use the single-file prescription without dummies. Co-Authored-By: Claude Opus 5.5 <[email protected]>
…tring contract admits Co-Authored-By: Claude Opus 5.5 <[email protected]> Claude-Session: https://claude.ai/code/session_01MFB6QPbxDLE71AUX5ieZ7n
This file contains hidden or bidirectional Unicode text that may be interpreted or compiled differently than what appears below. To review, open the file in an editor that reveals hidden Unicode characters.
Learn more about bidirectional Unicode characters
Sign up for free
to join this conversation on GitHub.
Already have an account?
Sign in to comment
Add this suggestion to a batch that can be applied as a single commit.This suggestion is invalid because no changes were made to the code.Suggestions cannot be applied while the pull request is closed.Suggestions cannot be applied while viewing a subset of changes.Only one suggestion per line can be applied in a batch.Add this suggestion to a batch that can be applied as a single commit.Applying suggestions on deleted lines is not supported.You must change the existing code in this line in order to create a valid suggestion.Outdated suggestions cannot be applied.This suggestion has been applied or marked resolved.Suggestions cannot be applied from pending reviews.Suggestions cannot be applied on multi-line comments.Suggestions cannot be applied while the pull request is queued to merge.Suggestion cannot be applied right now. Please check back later.
Requested by Matteo · project thread
Carries the 15 commits of #230 (the 7 T session's hot-fixes: bSSFP 2D bridged and gated trains with a flip-angle ramp catalyst, EPI FLEET calibration and ADC dwell, 2D fat saturation and saturation bands under a 40 T/m/s slew ceiling) plus one commit that makes its test suite pass:
bssfp2D_sequencereturns one sequence, or a chain of dummies and frame files for a free-running slice with dummies. Its return annotation is nowpp.Sequence | list[pp.Sequence], andtests/test_sequence_docstrings.pyadmits that third return kind (pypulseqpp.Sequence or list of pypulseqpp.Sequence, with an indented description) alongside a single sequence and a chain.pulserver #379 needs this on PyPI; it goes out as 0.0.64.